Research Article

Safety and Feasibility Using a Fluid-Filled Wire to Avoid Hydrostatic Errors in Physiological Intracoronary Measurements

Table 2

Height measurements and correction of sensor-tipped wire measurements (n = 19).

VesselHeight difference ± x (cm)Hydrostatic error (0.77 mmHg/cm) (mmHg)Pa at rest sensor-tipped (mmHg)Pd at rest sensor-tipped (mmHg)Pd at rest sensor-tipped after hydro correction (mmHg)Pa at hyperemia sensor-tipped (mmHg)Pd at hyperemia sensor-tipped (mmHg)Pd at hyperemia sensor-tipped after hydro correction (mmHg)

LAD−12.4−9.5958998.5978695.5
LAD−5.8−4.5988993.5846266.5
LAD−5.5−4.2837680.2806872.2
LAD−5.2−4.0124111115.01209296.0
LAD−4.9−3.8877376.8978588.8
LAD−4.3−3.3908689.3898285.3
LAD−4.2−3.2857982.2967477.2
LAD−4.0−3.1999396.1827174.1
LAD−4.0−3.1767174.1595053.1
LAD−3.8−2.9837072.9736163.9
LAD−3.7−2.81009496.8948486.8
LAD−3.6−2.8868082.8968284.8
LAD−2.4−1.8898384.8898485.8
LAD−1.2−0.9737070.9686565.9
RCA2.11.6113105103.4665250.4
LCX2.31.8828280.2867674.2
RCA3.02.3857875.7785855.7
RCA3.72.8868784.2948683.2
LCX3.93.0868380.0866663.0