Research Article

[Retracted] Acoustic Control through Gradient Coil Design Using a Finite-Difference-Based Method for MRI

Table 1

Characteristics of the designed coils.

Coil propertyOptimized coilConventional coil

Transport current (A)439.1447.0
Efficiency (μT·m−1·A−1)67.9166.74
Max field deviation, (%)5.05.0
Max ()/max () (%)3.03.0
Min. wire gap (mm)3.94.2
Inductance (μH)647574
Resistance (mΩ)474429
Displacement (RMS), f = 1750 Hz(μm)1.22.1
FOM1, η2/L (μ T2·m−2·A−2·H−1)7.137.76