Typical IL-13 calibration curve data obtained during assay validation.

IL-13 standard concentration (pg/mL)WellsODOD meanCV (%)BackCalc meanBackCalc   Cv in %Accuracy (%)Accuracy mean (%)

40A113.038 3.073 1.6 39.6 2.8697.05 99.05
A123.108101.06
30B112.543 2.592 2.6 30.2 3.997.78 100.56
B122.64103.33
20C111.897 1.926 2.1 20.2 2.6599.27 101.16
C121.955103.06
10D111.024 1.058 4.5 10.1 5.0497.07 100.66
D121.091104.25
5E110.556 0.559 0.5 4.9 0.6297.48 97.91
E120.56198.34
2.5F110.333 0.328 1.8 2.5 2.46102.45 100.7
F120.32498.94
1.25G110.21 0.213 1.5 1.3 2.9398.51 100.6
G120.215102.68
0.63H110.161 0.159 1.6 0.6 4.81104.68 101.23
H120.15897.79
0.31D90.135 0.137 2.2 0.3 11.35102.43 111.37
D100.139120.31
0.21E90.124 0.125 0.7 0.2 6.7583.33 87.51
E100.12691.69
0.16F90.123 0.123 0.5 0.1 5.8296.95 93.12
F100.12289.29