Development and Validation of an ELISA at Acidic pH for the Quantitative Determination of IL-13 in Human Plasma and Serum

Table 3

Parallelism assessment of IL-13 assay in human plasma under acidic conditions. The accuracy was calculated using the IL-13 concentration determined at dilution 1 : 2 as a reference value. Parallelism was considered acceptable when accuracy was between 70 and 130% for at least 5/6 samples at each dilution.

DilutionRaw result (pg/mL)Accuracy   ( / , %)
Observed ( )Expected ( )

Sample 1
1 : 20.425
1 : 30.2500.28388.2
1 : 40.1950.21391.8
1 : 6N.CN.C.N.C.

Sample 2
1 : 21.280
1 : 30.7600.85389.1
1 : 40.5250.64082.0
1 : 60.3620.42784.8

Sample 3
1 : 20.745
1 : 30.5170.497104.0
1 : 40.4250.373114.1
1 : 60.2850.248114.8

Sample 4
1 : 21.190
1 : 30.7630.79396.2
1 : 40.5400.59590.8
1 : 60.3650.39792.0

Sample 5
1 : 21.330
1 : 31.0200.887115.0
1 : 40.8300.665124.8
1 : 60.5530.443124.8

Sample 6
1 : 20.270
1 : 30.2070.180114.8
1 : 40.1480.135109.3
1 : 6N.CN.C.N.C.

N.C.: value not calculated as being outside the calibration range.