[Retracted] System Accuracy Evaluation of the GlucoRx Nexus Voice TD-4280 Blood Glucose Monitoring System

Table 2

Overview of the number of measurements obtained from the GlucoRx Nexus TD-4280 blood glucose monitoring device that were within the blood glucose concentration categories defined in both the ISO 15197:2003 and ISO 15197:2013 guidelines.
(a)

DIN EN ISO 15197:2003
Blood glucose concentration <75 mg/dL
(<4.2 mmol/L)
( = 5)
Blood glucose concentration ≥75 mg/dL
(≥4.2 mmol/L)
( = 100)

Blood glucose concentration categories defined in guidelines±5 mg/dL
(0.28 mmol/L)
±10 mg/dL
(0.55 mmol/L)
±15 mg/dL
(0.83 mmol/L)
±5%±10%±15%±20%
Number of results within category1/52/5 4/5 38/10077/100 87/100 93/100
Percentage of results
within category (%)
20408038778793

(b)

DIN EN ISO 15197:2013
Blood glucose concentration <100 mg/dL
(<5.55 mmol/L)
( = 34)
Blood glucose concentration ≥100 mg/dL
(≥5.55 mmol/L)
( = 71)

Blood glucose concentration categories defined in guidelines±5 mg/dL
(0.28 mmol/L)
±10 mg/dL
(0.55 mmol/L)
±15 mg/dL
(0.83 mmol/L)
±5%±10%±15%
Number of results within category22/3426/3431/3426/7159/7160/71
Percentage of results within category (%)64.776.591.236.683.184.5