Research Article

A Rational Adoption of the High Sensitive Assay for Cardiac Troponin I in Diagnostic Routine

Table 4

Sensitivity, specificity, and positive and negative predictive values (PPV, NPV) according to TnI and hsTnI results at T0 and percent delta changes from baseline (Δc) after 3 and 6 hours from baseline (T3 and T6).

SensitivitySpecificityPPVNPV

T0
 TnI71.0%86.9%39.0%96.2%
 hsTnI73.3%78.5%28.7%96.1%
T3
 ΔcTnI90.9%67.9%52.6%95.0%
 Δc hsTnI90.9%82.1%66.7%95.8%
T6
 ΔcTnI83.3%85.7%78.9%88.9%
 Δc hsTnI83.3%100%100%90.3%