Characteristics of Three Different Chemiluminescence Assays for Testing for SARS-CoV-2 Antibodies

Table 1

Multilevel likelihood ratios as well as sensitivity and specificity are given for the different tests at different cut-off levels (manufacturers’ cut-offs and half and double the manufacturers’ cut-offs) for ruling in or ruling out a COVID-19 diagnosis. +LR, positive likelihood ratio; -LR, negative likelihood ratio. CMIA, chemiluminescence microparticle immunoassay; LIA, luminescence immunoassay; ECLIA, electrochemiluminescence immunoassay; Sens, sensitivity; Spec, specificity; CI, confidence interval.

+LR at cut-off/2-LR at cut-off/2+LR at cut-off-LR at cut-off+LR at 2x cut-off-LR at 2x cut-off
(Spec, [95% CI])(Sens, [95% CI])(Spec, [95% CI])(Sens, [95% CI])(Spec, 95% CI)(Sens, 95% CI)

CMIA600.0433740.0564850.18
(98.4 [97.5,99.0])(95.8 [91.0-98.4])(99.8 [99.3,99.9])(94.4 [89.2,97.5])(99.8 [99.4,100])(81.7 [74.3,87.7])
LIA820.0353510.1210270.14
(98.8 [98.0,99.4])(96.5 [92.1,98.9])(99.8 [99.3 - 99.9])(88,2 [81.8,93.0])(100 [99.7,100])(86.1 [79.4 - 91.3])
ECLIA5750.0359580.042>9580.042
(99.8 [99.4,100])(96.5 [92.1,98.9])(99.9 [99.5,100])(95.8 [91.2,98.5])(100 [99.7,100])(95.8 [91.2,98.5])