Feasibility of Remote Real-Time Guidance of a Cardiac Examination Performed by Novices Using a Pocket-Sized Ultrasound Device

Table 1

Components of the CLUE Protocol.

SignsProbe siteDiagnostic targetsNormalAbnormal

LVSDParasternal long axisDistance from the anterior leaflet of the mitral valve to the septum during diastole ≤1 cm >1 cm
LAEParasternal long axisAnteroposterior diameter of the LA versus overlying aorta throughout the cardiac cycleAorta > LALA > Aorta
ULC+Bilateral lung apicesNumber of lung comets in either lung apex<3 ≥3
eCVPSubcostal Percent luminal diameter reduction of IVC with respiration, without forced “sniffing”>50%≤50%