Efficacy of Amflow®, a Real-Time-Portable Feedback Device for Delivering Appropriate Ventilation in Critically Ill Patients: A Randomised, Controlled, Cross-Over Simulation Study
Figure 4
Distribution bar plots for tidal volumes (ml) between the feedback group (bag-mask ventilation by using Amflow®) and the no-feedback group in three scenarios. (a) ARDS scenario (315–385 ml). (b) Cardiac arrest scenario (450–550 ml). (c) Head trauma scenario (630–770 ml).
(a) |
(b) |
(c) |