Research Article

A Validated RP HPLC-PAD Method for the Determination of Hederacoside C in Ivy-Thyme Cough Syrup

Table 4

Limits of detection (LOD) and quantitation (LOQ) of the HPLC method for the determination of hederacoside C in Ivy-Thyme cough syrup.

Standard deviation of the response ( ) (mg/mL) (mg/mL)

14831.80.0110.032

The value of SD was calculated as the standard deviation of intercepts of regression lines plotted, at the chromatographic responses of the calibration solutions (0.03–0.15 mg/mL), parallel to the calibration curve ( ).
For calculation of LOD and LOQ values, see Results and Discussion.