Research Article

Application of Certain π-Acceptors for the Spectrophotometric Determination of Alendronate Sodium in Pharmaceutical Bulk and Dosage Forms

Table 1

Selected spectral data for the determination of alendronate sodium by proposed spectrophotometric methods.

ParameterValues for
method A
Values for
method B

(nm)840465
Beer’s law verification range (μg mL−1)2–102–12
Molar absorptivity (L mole−1 cm−1)7.9 × 1044.6 × 104
Sandell’s sensitivity (μg ml−1 per 0.001 A)4.0 × 10−34.9 × 10−3
Regression equation (Y*)
Slope (b)0.02540.0351
Intercept (a)0.02380.0241
Correlation coefficient (r)0.99890.9997
Limit of detection (μg mL−1)0.190.30
Limit of quantification (μg mL−1)0.620.99

Y*: a + bC, where C is the concentration of analyte (μg/mL) and Y is absorbance unit.