Research Article

Determination of Diethyl Phthalate and Polyhexamethylene Guanidine in Surrogate Alcohol from Russia

Table 3

Precision and accuracy of the UV/VIS method for analyzing DEP in alcohol samples.

SampleWavelength (nm)RSD Intraday (%)
( )a
RSD Interday (%)
( )
Recovery (%) ( )

Authentic surrogate alcohol sample from Russia (1344 mg/L)2271.82.5
276 1.16.8

Authentic medicinal sample from Russia (311 mg/L)2270.90.9
2764.54.6

Authentic surrogate alcohol sample from Lithuania (608 mg/L)227(interference)
2761.93.0

Spiked vodka sample (500 mg/L)2271.01.2100
2765.56.6105

Spiked vodka sample (1000 mg/L)2271.31.695
2762.32.8105

aRSD: relative standard deviation.