Research Article
Determination of Total Selenium in Infant Formulas: Comparison of the Performance of FIA and MCFA Flow Systems
Table 1
Influence of the FIA system variables: multivariate design.
| Experiment | Flow rate HCl (mL min−1) | Flow rate NaBH4 (mL min−1) | Mixing coil length (cm) | Absorbance Se(IV) 50 μg · L−1 dissolution () |
| 1 | 1.3 | 1.7 | 150 | 0.044 | 2 | 2.9 | 3.6 | 150 | 0.069 | 3 | 3.5 | 4.5 | 50 | 0.076 | 4 | 1.3 | 1.7 | 100 | 0.047 | 5 | 2.9 | 3.6 | 100 | 0.069 | 6 | 2.2 | 2.8 | 100 | 0.062 | 7 | 3.5 | 4.5 | 100 | 0.072 |
|
|