Estimation of Measurement Uncertainties for the DGT Passive Sampler Used for Determination of Copper in Water

Table 6

Results from sensitivity analysis, showing the effect on total uncertainty of the passive sampler measurement from reductions in uncertainty of selected parameters.

ParameterChange in uncertaintyResult on total uncertainty

Effective area, 50% reductionReduction from 7.6% to 6.1% in overall relative uncertainty
Recovery factor, 50% reductionReduction from 7.6% to 6.9% in overall relative uncertainty
Diffusion boundary layer, From 0.05 mm to 0.014 mm standard uncertaintyReduction from 7.6% to 7.3% in overall relative uncertainty
Diffusion pathway thickness50% reductionReduction from 7.6% to 7.3% in overall relative uncertainty
Diffusion pathway thickness4 times increaseIncrease from 7.6% to 12.2% in overall relative uncertainty