An Optimized High Throughput Clean-Up Method Using Mixed-Mode SPE Plate for the Analysis of Free Arachidonic Acid in Plasma by LC-MS/MS
Table 2
Recoveries of arachidonic acid in spiked samples treated by four sample pretreatment methods.
| Concentration of arachidonic acid spiked in plasma | PPT ( = 5) | LLE ( = 5) | SPE ( = 5) | MAS-M ( = 5) | Recoveries (%) | RSD (%) | Recoveries (%) | RSD (%) | Recoveries (%) | RSD (%) | Recoveries (%) | RSD (%) |
| 500 ng/mL | 129.32 | 11.14 | 85.48 | 55.90 | 132.95 | 19.60 | 103.21 | 5.17 | 2 μg/mL | 130.42 | 2.06 | 67.05 | 82.21 | 94.66 | 9.21 | 99.38 | 5.34 |
|
|