Research Article

Stability-Indicating HPLC Determination of Gemcitabine in Pharmaceutical Formulations

Table 3

Precision study of the proposed method.

Concentration (µg/mL)Intraday precisionInterday precision
Conc. foundRSDConc. foundRSD
Mean ± SDMean ± SD

10.994 ± 0.0151.520.985 ± 0.0181.81
55.030 ± 0.0741.485.041 ± 0.0921.83
2019.920 ± 0.2431.2219.840 ± 0.2801.41
3029.886 ± 0.2920.9829.861 ± 0.2330.78
4544.759 ± 0.2550.5744.742 ± 0.3260.73