Research Article

Development and Validation of a GC-MS Method for the Detection and Quantification of Clotiapine in Blood and Urine Specimens and Application to a Postmortem Case

Table 1

Method validation parameters and data.

SelectivityLinearity ()Accuracy (bias%)Precision (RSD%)LODLLOQRecovery (%)
LowHighLowHighLowHigh

UrineNo interference signals0.999718−0.6143.21.2 ng/mL3.9 ng/mL10399

BloodNo interference signals0.9996−18−0.5193.01.3 ng/mL4.3 ng/mL9395