Research Article

Determination of Chlorpromazine, Haloperidol, Levomepromazine, Olanzapine, Risperidone, and Sulpiride in Human Plasma by Liquid Chromatography/Tandem Mass Spectrometry (LC-MS/MS)

Table 1

Optimal instrumental settings.

Injection

Injection volume20 μl

Injection temperature40°C

Flush volume250 μl

Pre-inject Flush2

Post-inject Flush2

Flush speedMedium

Scan

Type of scanMRM: Multiple reaction monitoring

Mode of scanPositive

Source/Gas parameter

TEM (temperature)500 K

GS1 (gas source 1)40 psi

GS2 (gas source 2)50 psi

CUR (curtain gas)10 psi

CAD (collision gas)Medium

IS (ion spray voltage)5,000 v

IHE (interface heater)On