Research Article

Determination of Trace Antimony (III) in Water Samples with Single Drop Microextraction Using BPHA-[C4mim][PF6] System Followed by Graphite Furnace Atomic Absorption Spectrometry

Table 4

Determination of Sb (III) in water samples.

SamplesAdded (μg·L−1)Found (μg·L−1)Recovery (%)

Bottle mineral water0< Limits of detection-
0.10.102 ± 0.01102±1
0.40.401 ± 0.008100±2

River water0< Limits of detection-
0.10.104 ± 0.015104±2
0.40.407 ± 0.012102±3

Tap water0< Limits of detection-
0.10.098 ± 0.01198±1
0.40.398 ± 0.00999±2