Operating conditions for the ICP-MS instrument.

ICP-MS: Agilent 7700x
 Plasma conditions:
  RF power1.55 kW
  Plasma gas flow rate15.0 L min−1 Ar
  Auxiliary gas flow rate0.90 L min−1 Ar
  Makeup gas flow rate0 L min−1 Ar
  Carrier gas flow rate1.05 L min−1 Ar
  Sampling depth (mm from load coil)8.0 mm
  Cell gasHe mode:4.3 mL min−1
H2 mode:6.0 mL min−1
  NebulizerMicro Mist
  Sample uptake rate0.45 mL min−1

 Data acquisition:
  Accumulation time0.3-1.0 s / point
  Data point3 points / peak
  Repetition3 times