Results of capillary electrophoresis method validation and percentage content determination.
PARAMETERS
RESULTS
Content of ERTM
99.94%
Linearity
y = 2281.7x – 24495 R2 = 0.9994 (70 to 120 µg mL−1)
Repeatability
RSD = 1.62%
Intermediate precision
1st day 102.84%; 2nd day 99.83% and 3rd day 99.15% - RSD = 0.85%