Research Article

Application of the Principles of Green Chemistry for the Development of a New and Sensitive Method for Analysis of Ertapenem Sodium by Capillary Electrophoresis

Table 4

Results of capillary electrophoresis method validation and percentage content determination.

PARAMETERSRESULTS

Content of ERTM99.94%
Linearityy = 2281.7x – 24495 R2 = 0.9994 (70 to 120 µg mL−1)
RepeatabilityRSD = 1.62%
Intermediate precision1st day 102.84%; 2nd day 99.83% and 3rd day 99.15% - RSD = 0.85%
Accuracy100.59%, RSD = 1.09%
LOD0.77 µg mL−1
LOQ2.32 µg mL−1
Recovery100.59%

determinations; of three determinations.