Analysis of Acrylamide in Dried Blood Spots of Lung Cancer Patients by Ultrahigh-Performance Liquid Chromatography Tandem Mass Spectrometry
Table 2
Summary of the full validation result.
Analyte
QC
Conc. (μg/mL)
Precision (%CV)
Accuracy (%bias)
Matrix effect
Internal standard-normalized matrix factor
Within run
Between run
Within run
Between run
Mean % ± SD
%CV
Mean % ± SD
%CV
AA
LLOQ
2.50
7.07
9.51
−10.35
−11.73
QCL
7.50
3.89
4.12
6.12
9.88
105.68 ± 12.34
11.68
0.91 ± 0.13
14.11
QCM
50.0
3.12
4.48
9.49
13.71
QCH
75.00
1.89
5.41
−6.29
11.75
96.33 ± 8.77
9.11
0.98 ± 0.10
10.11
Propranolol
117.05 ± 5.77
4.93
QC: quality control, LLOQ: lower limit of quantification, QCL: low level, QCM: medium level, QCH: high level, and CV: variability, number of experiments.