Fiberoptic-Coupled Spectrofluorometer with Array Detection as a Process Analytical Chemistry Tool for Continuous Flow Monitoring of Fluoroquinolone Antibiotics
Table 1
The analytical performance of the method.
Parameter
Moxifloxacin
Levofloxacin
Ofloxacin
Ciprofloxacin
λ
max
(nm)
525
495
497
489
Linear range (
μ
g·mL
−1
)
2–20
0.25–12
0.25–10
1–8
Relative standard deviation
(
n
= 7%)
0.527
0.517
0.468
0.561
Detection limit (
μ
g·mL
−1
)
0.093
0.036
0.035
0.081
LOQ (
μ
g·mL
−1
)
0.308
0.1185
0.1155
0.269
Correlation coefficient
0.9979
0.9972
0.9999
0.9992
Concentration: 5
μ
g·mL
−1
.