Research Article

A Validated RP-HPLC Stability Method for the Estimation of Chlorthalidone and Its Process-Related Impurities in an API and Tablet Formulation

Table 3

Peak results for the spiked solution containing the chlorthalidone API and in-house process-related impurities.

Sample nameAreaRTRRTResolutionTheoretical platesTailing factorPurity anglePurity threshold

Chlorthalidone551505456.729ā€”2.5442401.219.7861.001
Stage II93895633.0564.91223.10475071.260.1831.129
Stage III7271299.7831.4533.2067701.170.1801.108