Gang Peng, Yadong Yu, Xiaojun Chen, He Huang, "Highly Sensitive Amperometric α-Ketoglutarate Biosensor Based on Reduced Graphene Oxide-Gold Nanocomposites", International Journal of Analytical Chemistry, vol. 2020, Article ID 4901761, 10 pages, 2020. https://doi.org/10.1155/2020/4901761

Highly Sensitive Amperometric α-Ketoglutarate Biosensor Based on Reduced Graphene Oxide-Gold Nanocomposites

Academic Editor: Victoria F. Samanidou
Received22 Jan 2020
Revised31 May 2020
Accepted15 Jun 2020
Published01 Aug 2020

Abstract

Herein, a rapid and highly sensitive amperometric biosensor for the detection of α-ketoglutarate (α-KG) was constructed via an electrochemical approach, in which the glutamate dehydrogenase (GLUD) was modified on the surface of reduced graphene oxide-gold nanoparticle composite (rGO-Aunano composite). The rGO-Aunano composite was one-step electrodeposited onto glassy carbon electrode (GCE) surface and was characterized by scanning electron microscopy (SEM), energy dispersive X-ray spectroscopy (EDS), and electrochemical techniques. In addition, the rGO-Aunano/GCE was also found to electrocatalyze the oxidation of β-nicotinamide adenine dinucleotide (NADH) at the peak potential of 0.3 V, which was negatively shifted compared with that at bare GCE or Aunano/GCE, illustrating better catalytic performance of rGO-Aunano. After the modification of GLUD, the GLUD/rGO-Aunano/GCE led to effective amperometric detection of α-KG through monitoring the NADH consumption and displayed a linear response in the range of 66.7 and 494.5 μM, with the detection limit of 9.2 μM. Moreover, the prepared GLUD/rGO-Aunano/GCE was further evaluated to be highly selective and used to test α-KG in human serum samples. The recovery and the RSD values were calculated in the range of 97.9–102.4% and 3.8–4.5%, respectively, showing a great prospect for its real application.

1. Introduction

α-Ketoglutarate (α-KG) is an important metabolic marker of early diagnosis of cancer and microbial fermentation monitoring and also is a vital intermediate in the tricarboxylic acid cycle as well as a node to connect the carbon-nitrogen metabolism in cells [14]. It has been emphasized to possess exciting angiogenesis suppressor activity [57]. Furthermore, micronutrient application of α-KG has exhibited beneficial effects on several malignant tumors with only minor negative effects on normal cells [4,5]. Therefore, accurate and sensitive methods for the detection of α-KG are urgently required. Till now, several methods have been developed for the quantification of α-KG, such as electrochemical [8], gas chromatography-mass spectrometry (GC-MS) [4], and high-performance liquid chromatography (HPLC) [5, 9]. Among them, electrochemical methods have become considerably meaningful due to the advantages of convenience, high speed, low cost, and easy-to-use. Currently, Wang and his coworkers have developed an electrochemical biosensor for rapid and sensitive detection of α-KG based on ruthenium-rhodium modified carbon fiber enzyme microelectrode [8]. However, a large overpotential was found at the microelectrode and the detection of α-KG was prone to suffer interference from other compounds. Therefore, it is important to prepare a capable modified electrode with lower overpotential for α-KG detection and meanwhile maintain the bioactivity of the immobilized enzyme.

Graphene (Gr) has become extremely attractive in many fields since it has superiorities such as good biocompatibility [10], large surface-to-volume ratio, excellent conductivity [11], electron mobility, and flexibility [12]. In the electrochemical biosensing fields, its large surface-to-volume ratio is helpful for increasing the loading amount of the enzyme, and its excellent conductivity is favourable for transferring electrons between the electrode surface and electrolyte [1319]. However, the practical applications of Gr are challenged by its irreversible agglomeration or even restack to form graphite due to the Van der Waals interactions in the drying state [20]. Recent research found that noble metal nanoparticles could be used to effectively prevent the massive agglomeration of Gr sheets. Si got a Pt nanoparticles (Ptnano)-Gr composite with the Gr partially exfoliated from its drying aqueous dispersions [21]. Tang et al. founded a rapid and efficient one-step approach to prepare Gr-Agnano composite by simultaneous reduction of Gr oxide (GO) and Ag+ with formaldehyde as the reducing agent [22]. Li et al. developed a sensor for the detection of paracetamol based on Pdnano-GO composite, which was obtained by a one-pot chemical reduction using Pd2+ as a precursor without large aggregation [23]. Also, noble metal nanoparticles-Gr composites have been found to have excellent catalytic performance in electrochemical biosensors [2426]. Leng et al. demonstrated that the Pdnano-GO composite could electrocatalyze the oxidation of rutin, with the detection limit of 0.001 μM [27]. Govindhan et al. prepared Aunano-Gr composite which could catalyze NADH with the detection limit as low as 1.13 nM, since the composite provided a large electrochemical active surface area and a favourable environment for electron transfer from NADH to the electrode [28]. Therefore, in this work, we were ready to construct an α-KG biosensor based on reduced GO-Aunano (rGO-Aunano) composite.

In this paper, the rGO-Aunano composite was one-step electrodeposited onto the glassy carbon electrode (GCE) surface, which exhibited a lower working potential for the electrochemical oxidation of NADH, and served as the platform for the immobilization of glutamate dehydrogenase (GLUD). α-KG was detected via monitoring the NADH consumption, and the mechanism was described as in the following equation [8]:

Catalyzed by GLUD, α-KG was conversed to L-glutamate in the presence of NH4+ and NADH. The more amount of α-KG contained in solution, the more NADH was consumed. Then, if a certain amount of NADH was added in advance, the remaining amount of NADH would decrease along with the increase of α-KG. Therefore, the concentration of α-KG was inversely proportional to the catalytic current of NADH, which provided the quantitative basis for α-KG detection.

2. Materials and Methods

2.1. Materials and Reagents

GO was purchased from Suzhou Hengqiu Co., Ltd. (Suzhou, China). Chloroauric acid and ammonium chloride were obtained from Sinopharm Chemical Reagent Co., Ltd. (Shanghai, China). β-nicotinamide adenine dinucleotide reduced dipotassium salt (NADH), L-Glutamic dehydrogenase from bovine liver-Type III (GLUD), and α-ketoglutaric acid sodium salt (α-KG) were supplied by Sigma Chemical Co., Ltd. (St. Louis, MO, USA). The human serum samples were provided by Jiangsu Center for Clinical Laboratory (JSCCL). All other chemicals were of analytical grade and used without further purification. 0.05 M pH 9.0 of carbonate buffer solution (CBS) and 0.1 M pH 7.2 of phosphate buffered solution (PBS) were used as the electrolyte. Double-distilled water was used throughout the experiments.

2.2. Characterization and Electrochemical Measurement

Surface morphologies of rGO and rGO-Aunano films were investigated by scanning electron microscopy (SEM, JEOL, JSM-6510, Japan). The elemental composition analysis was performed by energy dispersive X-ray spectroscopy (EDS, Vantage 4105, NORAN). All electrochemical measurements were carried out on a CHI 650E electrochemical workstation (Shanghai Chenhua Instrument Company, China), employing a typical three-electrode cell system. A modified or bare GCE was utilized as the working electrode (Ф = 3 mm), whereas a saturated calomel electrode (SCE) as a reference and a platinum foil electrode as a counter. All potentials were measured against SCE. All experiments were purged with high-purity nitrogen to remove oxygen and done at room temperature (∼25°C).

2.3. Fabrication of α-KG Electrochemical Biosensor

The fabrication process of the electrochemical biosensor was illustrated in Scheme 1. Prior to electrodeposition, the GCE was polished using 0.3 and 0.05 μm alumina powder until a mirror-shiny surface was obtained, and followed by ultrasonication in the ethanol and double-distilled water for 2 min, respectively. The cleaned GCE was then modified by the electrochemical codeposition of the GO and Aunano. According to the method reported by Liu et al. [29], GO was exfoliated in CBS by ultrasonication for 20 min to form a homogeneous few-layer GO dispersion. Cyclic voltammetric (CV) reduction was performed in the deposition solutions containing 1.0 mg·mL−1 of GO and 200 μM of chloroauric acid with magnetic stirring. The CV was carried out between −1.4 and 0.6 V at a rate of 20 mV·s−1, and the deposition amount was optimized as six potential cycles. Then, 8 μL of 110 kU·L−1 GLUD PBS (0.1 M, pH 7.2) was dropped onto rGO-Aunano/GCE surface. After natural drying, it was dipped into a 2 wt.% of glutaraldehyde for 3 s to form a protective film and stored in PBS at 4°C before use.

3. Results and Discussion

3.1. Characterization of rGO-Aunano Composite

CV was recorded during the electrodeposition of GO on GCE (Figure 1(a)). There is one anodic peak (I) and two cathodic peaks (II and III), corresponding to the redox pair of some oxygen-containing groups [30] and the irreversible electrochemical reduction of GO [31], respectively. The on-going increase of the peak current with successive potential sweeps indicated that the rGO from GO dispersion was successfully deposited onto GCE. Figure 1(b) shows the CV of codeposition of GO and HAuCl4, with a completely different appearance from Figure 1(a). The reduction current was larger than that of GO electrolysis, indicating that the deposition of rGO and Aunano onto the surface of GCE was achieved [29]. The number of electrodeposition cycles was also optimized since the thickness of the rGO-Aunano composite film would influence the conductivity and stability of the rGO-Aunano/GCE. This part would be interpreted in detail in the section of the optimization of the experimental parameters.

The morphologies of rGO and rGO-Aunano were characterized by SEM. Figure 1(c) shows the SEM image of rGO film, in which a typically wrinkled texture was displayed. In Figure 1(d), you could find that Aunano uniformly scattered over rGO to make the rGO-Aunano composite. The Aunano prevented the agglomeration of rGO; meanwhile, rGO enhanced the dispersion of Aunano, both of which improve the conductivity and stability of rGO-Aunano composite film. Further, the EDS spectrum confirmed the elemental composition of rGO-Aunano, in which Si element comes from the Si substrate. The FT-IR spectra of GO and rGO are shown in Figure 2. The FT-IR spectrum of GO (a) exhibits the characteristic absorptions from oxygen-containing functional groups. In detail, the absorption band at 3437 and 1397 cm−1 can be assigned to the stretching vibration and deformation vibration of O-H, respectively. The band at 1052 cm−1 belongs to the C-O (alkoxy), while the band at 1633 cm−1 derives from the vibration of the adsorbed water molecules and/or the contribution of the skeletal vibration of unoxidized graphitic domains. After the reduction, the peaks at 1070 cm−1 assigned to epoxy groups are decreased significantly, clearly indicating the removal of oxygen-containing groups of GO (curve b in Figure 2) and suggesting the successful reduction of GO [32].

3.2. Electrochemical Performance of rGO-Aunano Composite

Figure 3(a) presents the CVs of four different modified GCEs in 0.1 M KCl containing 5 mM of [Fe(CN)6]3−/4− at a scan rate of 50 mV·s−1. The current response of rGO/GCE (curve B towards [Fe(CN)6]3−/4−) was larger than that of bare GCE (curve A), owing to that rGO enhanced the electron transfer ability. The similarly enhanced peak current at Aunano/GCE (curve C) attributed to the good conductivity of Aunano. After the rGO-Aunano composite modified onto the GCE surface (curve D), the peak current rose further owing to the synergy from rGO and Aunano. The electrochemical properties of these modified electrodes were also characterized by EIS. Figure 3(b) shows nearly straight lines at rGO/GCE (curve B), Aunano/GCE (curve C), and rGO-Aunano/GCE (curve D), revealing rapid electron transfer between the electrode surface and the electrolyte. Only a small semicircle was observed at bare GCE (curve A), which meant about 100 Ω of electron transfer resistance in this system.

CV technique was also performed to evaluate the electrocatalytic oxidation effect of NADH on different electrodes in 0.1 M PBS (pH 7.2) containing 0.5 mM of NADH at a scan rate of 50 mV·s−1. As shown in Figure 3(c), a very small anodic peak was obtained at 0.7 V at the bare GCE (curve A), which meant that the direct oxidation of NADH at the bare electrode was difficult. After the modification of Aunano, the anodic peak increased and the working potential negatively shifted to 0.62 V, which might be owing to the good oxidation catalytic ability of Aunano (curve B) [32]. For the rGO/GCE as curve C, a small oxidation peak at 0.24 V was found, showing the excellent catalytic capacity of rGO to NADH and the activation energy of NADH oxidation was declined on the surface of rGO. The reason might be that the presence of abundant defects and edge plane graphite structures in rGO was thought to facilitate the heterogeneous charge transfer at the electrode interface [25, 33, 34]. Thus, the electrons released by NADH could be transferred swiftly with the aid of rGO. After the combination of Aunano and rGO, the oxidation peak current of NADH was similar to that of Aunano/GCE, and the peak potential appeared at 0.32 V. The synergic effect of the rGO and Aunano exhibited the capability as a powerful catalyst to the oxidation of NADH [35]. In addition, we also found that no obvious peaks were observed on all these four modified electrodes in PBS without NADH (Figure S1), indicating NADH could be catalytically oxidized with the aid of Aunano and rGO.

The effect of the scan rate on the oxidation current of NADH was also investigated. The CVs of the rGO-Aunano/GCE at different scan rates were recorded as Figure 3(d). A good linear relationship of the anodic peak current with the square root of the scan rate in the range of 30 and 100 mV·s−1 was presented as the inset of Figure 3(d). The result revealed the oxidation of NADH on the rGO-Aunano/GCE was a diffusion-controlling process, owing to the fast electron transfer rate between rGO-Aunano and the electrolyte.

3.3. Optimization of the Experimental Parameters

To improve the performance of the NADH sensor, the effect of determination conditions such as the electrodeposition cycles, the applied potential, and the solution pH have been investigated in detail.

The number of the electrodeposition cycles was optimized by measuring the catalytic current to 0.5 mM NADH using rGO-Aunano/GCE with a different rGO-Aunano film. The more electrodeposition cycles exerted, the thicker the rGO-Aunano composite film would be obtained. As shown in Figure 4(a), the oxidation peak current increased sharply as increasing the electrodeposition cycle number from 2 to 6, due to the enhanced conductivity and catalytic activity along with increasing rGO-Aunano deposition. However, the peak current decreased when the number of electrodeposition cycles was more than 6, owing to the fact that the thicker rGO-Aunano film was inclined to drop off. Thus, 6 cycles of electrodeposition of rGO-Aunano deposition were selected in this work.

In order to achieve the best electrocatalytic effect while reducing the overpotential, the applied potential should be optimized. The effect of the applied potential was shown in Figure 4(b). As the applied potential increased from 0.1 to 0.3 V, the peak current increased gradually. Further, increasing the applied potential from 0.3 to 0.7 V led to a relatively constant current value. Therefore, 0.3 V was selected as the optimal applied potential.

The effect of the solution pH on the oxidation of NADH at rGO-Aunano/GCE was also studied by monitoring the peak current of NADH in 0.1 M PBS with different pH values from 6.0 to 8.0. Figure 4(c) shows that the oxidation peak current of NADH increased along with the solution pH value from 6.0 to 7.2, and then decreased. Therefore, pH 7.2 was selected as the optimum for further studies.

3.4. Detection of α-KG Using GLUD/rGO-Aunano/GCE

Based on the catalytic mechanism described in the introduction part, the concentration of α-KG could be determined by the depleted amount of NADH at a GLUD/rGO-Aunano/GCE. That is to say, the content of α-KG was inversely proportional to the catalytic current of NADH. To obtain the optimal performance of the biosensor, the concentration of the immobilized GLUD needed to be investigated further. As seen from Figure 5(a), the current response of NADH decreased quickly along with the GLUD concentration increased from 27.5 to 110 kU·L−1, showing a fast transform from α-KG to L-glutamate. Then, the current tended to become steady with a further increase in GLUD concentration, owing to the saturation of GLUD loading capacity. From an economic perspective, 110 kU·L−1 of GLUD was selected in our experiments.

Figure 5(b) shows a typical amperometric current-time curve recorded at GLUD/rGO-Aunano/GCE for the successive additions of 66.7 μM α-KG in a stirred pH 7.2 PBS containing 1 mM NADH and 1 mM NH4Cl. The response was very fast and the steady-state current response was attained in less than 5 s. The linear response range at GLUD/rGO-Aunano/GCE was from 66.7 to 494.5 μM, with the sensitivity of 454 μA M−1 and a correlation coefficient of 0.9992. The calculated limit of detection was found to be 9.2 μM (S/N = 3). After the GLUD immobilization, the biosensor sensitivity was decreased due to the enzyme layer act as a barrier that hindered NADH transport. To our knowledge, there are few reports on electrochemical biosensors for the measurement of α-KG, and nearly a few references could be compared. As listed in Table 1, our sensing system exhibited a comparable linear range with a bienzymatic flow injection system [37] and a similar detection limit with the other electrochemical biosensors, suggesting the proposed sensor had a good performance in α-KG detection.


α-KG sensorLinear range (μM)Detection limit (μM)Ref.

GLUD-Ru/Rh100–60020[8]
GLUD-rMNs-Aunano11.12–52.946.25[36]
GDH-GlOx0–5007[37]
GLUD/rGO-Aunano66.7–494.59.2This work

3.5. Reproducibility, Stability, and Selectivity of the Biosensor

The reproducibility of GLUD/rGO-Aunano/GCE was investigated in 0.1 M PBS containing 1 mM NADH and 1 mM NH4Cl via recoding the current response during the successive addition of 66.7 μM of α-KG. Six sensors were prepared in different batches, and 10 successive measurements were implemented for each electrode, and the RSD of the current response was 4.8 and 4.2%, respectively. The stability of GLUD/rGO-Aunano/GCE was also evaluated. The current response for 312.5 μM α-KG was measured, and there was nearly no apparent loss during ∼1000 s’ operation, as shown in Figure 6(a). Such good stability of GLUD/rGO-Aunano/GCE was attributed to the protection of the glutaraldehyde membrane and the biocompatibility of rGO-Aunano for GLUD immobilization.

Selectivity is important for the practical application of biosensors. An assessment of the interference on the amperometric response to NADH was examined at the GLUD/rGO-Aunano/GCE in the presence of other oxidizable substances, such as dopamine (DA), ascorbic acid (AA), and uric acid (UA). Figure 6(b) shows the amperometric responses at GLUD/rGO-Aunano/GCE after the addition of various interferents. In the presence of 0.25 mM NADH, 0.1 mM DA or UA could not induce apparent interference, while 0.1 mM AA generated an obvious interference, since AA could be oxidized at lower potential compared with DA and UA [38, 39]. Thus, AA should be removed when the biosensor was used for practical application. One possible method for removing AA was the immobilization of ascorbate oxidase onto the electrode surface [40].

3.6. Real Sample Analysis in Human Serum

The designed GLUD/rGO-Aunano/GCE was further evaluated in real human serum using the standard addition method. The human serum was diluted 10 times with 0.1 M pH 7.2 PBS. Then, the solution was transferred into the electrochemical cell for analysis. The amperometric current-time measurement was employed for the recovery test to determine α-KG in the human serum samples. As listed in Table 2, the recoveries of α-KG between 97.9 and 102.4% with RSD values in the range of 3.8 to 4.5% were obtained, indicating the strong potential application prospect in clinical diagnosis.


SampleAdded (μM)Found (μM)aRecovery (%)RSD (%)

1100102.4102.44.5
2200197.698.83.8
3300293.797.94.3

aAverage of three measurements.

4. Conclusion

In this work, an α-KG biosensor based on the rGO-Aunano film was successfully prepared. The rGO-Aunano/GCE platform resulted in the improvement of electrocatalytic activity towards the oxidation of NADH with lower working potential, higher sensitivity, stability, and reproducibility. After the modification of GLUD, the electrode exhibited a sensitive response to α-KG in PBS via the consumption of NADH, implying good stability of GLUD on biocompatible rGO-Aunano platform and rapid electron transfer ability between electrode and α-KG. Moreover, the biosensor was applied to determine α-KG in human serum with satisfactory recoveries, illustrating its good selectivity and anti-interference. To summarize, the α-KG sensing strategy reported in this work could be expected for real applications in the future.

Data Availability

The data that support the findings of this study are available from the corresponding author upon reasonable request.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Acknowledgments

This study was financially supported by the National Natural Science Foundation of China (21575064), the Six Talent Peaks Project in Jiangsu Province (2016-SWYY-022), the Qinglan Project of Jiangsu Education Department (2016), and the Program for Innovative Research Team in Universities of Jiangsu Province (2015).

Supplementary Materials

Figure S1: CVs of different working electrodes in 0.1 M pH 7.2 PBS without NADH at a scan rate of 50 mV·s−1. (Supplementary Materials)

References

  1. S. Zhao, Y. Lin, W. Xu et al., “Glioma-derived mutations in IDH1 dominantly inhibit IDH1 catalytic activity and induce HIF-1,” Science, vol. 324, no. 5924, pp. 261–265, 2009. View at: Publisher Site | Google Scholar
  2. H. A. Krebs and P. P. Cohen, “Metabolism of α-ketoglutaric acid in animal tissues,” Biochemical Journal, vol. 33, no. 11, pp. 1895–1899, 1939. View at: Publisher Site | Google Scholar
  3. J. E. Baldwin and H. Krebs, “The evolution of metabolic cycles,” Nature, vol. 291, no. 5814, pp. 381-382, 1981. View at: Publisher Site | Google Scholar
  4. B. M. Wagner, F. Donnarumma, R. Wintersteiger, W. Windischhofer, and H. J. Leis, “Simultaneous quantitative determination of α-ketoglutaric acid and 5-hydroxymethylfurfural in human plasma by gas chromatography-mass spectrometry,” Analytical and Bioanalytical Chemistry, vol. 396, no. 7, pp. 2629–2637, 2010. View at: Publisher Site | Google Scholar
  5. K. Michail, H. Juan, A. Maier, V. Matzi, J. Greilberger, and R. Wintersteiger, “Development and validation of a liquid chromatographic method for the determination of hydroxymethylfurfural and alpha-ketoglutaric acid in human plasma,” Analytica Chimica Acta, vol. 581, no. 2, pp. 287–297, 2007. View at: Publisher Site | Google Scholar
  6. W. Schreibmayer, R. J. Schaur, H. M. Tillian, E. Schauenstein, and K. Hagmüller, “Tumor host relations,” Journal of Cancer Research and Clinical Oncology, vol. 97, no. 2, pp. 137–144, 1980. View at: Publisher Site | Google Scholar
  7. R. J. Schaur, W. Schreibmayer, H.-J. Semmelrock, H. M. Tillian, and E. Schauenstein, “Tumor host relations,” Journal of Cancer Research and Clinical Oncology, vol. 93, no. 3, pp. 293–300, 1979. View at: Publisher Site | Google Scholar
  8. S. Poorahong, P. Santhosh, G. V. Ramírez et al., “Development of amperometric α-ketoglutarate biosensor based on ruthenium-rhodium modified carbon fiber enzyme microelectrode,” Biosensors and Bioelectronics, vol. 26, no. 8, pp. 3670–3673, 2011. View at: Publisher Site | Google Scholar
  9. H. Terada, T. Hayashi, S. Kawai, and T. Ohno, “High-performance liquid chromatographic determination of pyruvic acid and α-ketoglutaric acid in serum,” Journal of Chromatography A, vol. 130, pp. 281–286, 1977. View at: Publisher Site | Google Scholar
  10. X. Y. Zhang, J. L. Yin, C. Peng et al., “Distribution and biocompatibility studies of graphene oxide in mice after intravenous administration,” Carbon, vol. 49, no. 3, pp. 986–995, 2011. View at: Publisher Site | Google Scholar
  11. A. A. Balandin, S. Ghosh, W. Bao et al., “Superior thermal conductivity of single-layer graphene,” Nano Letters, vol. 8, no. 3, pp. 902–907, 2008. View at: Publisher Site | Google Scholar
  12. U. Stöberl, U. Wurstbauer, W. Wegscheider, D. Weiss, and J. Eroms, “Morphology and flexibility of graphene and few-layer graphene on various substrates,” Applied Physics Letters, vol. 93, no. 5, Article ID 051906, 2008. View at: Publisher Site | Google Scholar
  13. Y. Liu, D. Yu, C. Zeng, Z. Miao, and L. Dai, “Biocompatible graphene oxide-based glucose biosensors,” Langmuir, vol. 26, no. 9, pp. 6158–6160, 2010. View at: Publisher Site | Google Scholar
  14. K. Zhou, Y. Zhu, X. Yang, J. Luo, C. Li, and S. Luan, “A novel hydrogen peroxide biosensor based on Au-graphene-HRP-chitosan biocomposites,” Electrochimica Acta, vol. 55, no. 9, pp. 3055–3060, 2010. View at: Publisher Site | Google Scholar
  15. J. Yang, S. Deng, J. Lei, H. Ju, and S. Gunasekaran, “Electrochemical synthesis of reduced graphene sheet-AuPd alloy nanoparticle composites for enzymatic biosensing,” Biosensors and Bioelectronics, vol. 29, no. 1, pp. 159–166, 2011. View at: Publisher Site | Google Scholar
  16. Q. He, J. Liu, X. Liu et al., “A promising sensing platform toward dopamine using MnO2 nanowires/electro-reduced graphene oxide composites,” Electrochimica Acta, vol. 296, pp. 683–692, 2019. View at: Publisher Site | Google Scholar
  17. G. Li, Y. Xia, Y. Tian et al., “Review-recent developments on graphene-based electrochemical sensors toward nitrite,” Journal of The Electrochemical Society, vol. 166, no. 12, pp. B881–B895, 2019. View at: Publisher Site | Google Scholar
  18. Q. Li, Y. Xia, X. Wan et al., “Morphology-dependent MnO2/nitrogen-doped graphene nanocomposites for simultaneous detection of trace dopamine and uric acid,” Materials Science and Engineering: C, vol. 109, Article ID 110615, 2020. View at: Publisher Site | Google Scholar
  19. G. Li, J. Wu, H. Jin et al., “Titania/electro-reduced graphene oxide nanohybrid as an efficient electrochemical sensor for the determination of allura red,” Nanomaterials, vol. 10, no. 2, p. 307, 2020. View at: Publisher Site | Google Scholar
  20. D. Li, M. B. Müller, S. Gilje, R. B. Kaner, and G. G. Wallace, “Processable aqueous dispersions of graphene nanosheets,” Nature Nanotechnology, vol. 3, no. 2, pp. 101–105, 2008. View at: Publisher Site | Google Scholar
  21. Y. Si and E. T. Samulski, “Exfoliated graphene separated by platinum nanoparticles,” Chemistry of Materials, vol. 20, no. 21, pp. 6792–6797, 2008. View at: Publisher Site | Google Scholar
  22. X.-Z. Tang, Z. Cao, H.-B. Zhang, J. Liu, and Z.-Z. Yu, “Growth of silver nanocrystals on graphene by simultaneous reduction of graphene oxide and silver ions with a rapid and efficient one-step approach,” Chemical Communications, vol. 47, no. 11, pp. 3084–3086, 2011. View at: Publisher Site | Google Scholar
  23. J. Li, J. Liu, G. Tan et al., “High-sensitivity paracetamol sensor based on Pd/graphene oxide nanocomposite as an enhanced electrochemical sensing platform,” Biosensors and Bioelectronics, vol. 54, pp. 468–475, 2014. View at: Publisher Site | Google Scholar
  24. L. Zhang, L. Yang, L. Zhang, and D. W. Li, “Electrocatalytic oxidation of NADH on graphene oxide and reduced graphene oxide modified screen-printed electrode,” International Journal of Electrochemical Science, vol. 6, pp. 819–829, 2011. View at: Google Scholar
  25. H. Chang, X. Wu, C. Wu, Y. Chen, H. Jiang, and X. Wang, “Catalytic oxidation and determination of β-NADH using self-assembly hybrid of gold nanoparticles and graphene,” The Analyst, vol. 136, no. 13, pp. 2735–2740, 2011. View at: Publisher Site | Google Scholar
  26. P. K. Simson, R. Manjunatha, C. Nethravathi, G. S. Suresh, M. Rajamathi, and T. V. Venkatesha, “Electrocatalytic oxidation of NADH on functionalized graphene modified graphite electrode,” Electroanalysis, vol. 23, pp. 842–849, 2011. View at: Publisher Site | Google Scholar
  27. J. Leng, P. Li, L. Bai, Y. Peng, Y. Yu, and L. Lu, “Facile Synthesis of Pd nanoparticles-graphene oxide hybrid and its application to the electrochemical determination of rutin,” International Journal of Electrochemical Science, vol. 10, pp. 8522–8530, 2015. View at: Google Scholar
  28. M. Govindhan, M. Amiri, and A. Chen, “Au nanoparticle/graphene nanocomposite as a platform for the sensitive detection of NADH in human urine,” Biosensors and Bioelectronics, vol. 66, pp. 474–480, 2015. View at: Publisher Site | Google Scholar
  29. C. Liu, K. Wang, S. Luo, Y. Tang, and L. Chen, “Direct electrodeposition of graphene enabling the one-step synthesis of graphene-metal nanocomposite films,” Small, vol. 7, no. 9, pp. 1203–1206, 2011. View at: Publisher Site | Google Scholar
  30. L. Chen, Y. Tang, K. Wang, C. Liu, and S. Luo, “Direct electrodeposition of reduced graphene oxide on glassy carbon electrode and its electrochemical application,” Electrochemistry Communications, vol. 13, no. 2, pp. 133–137, 2011. View at: Publisher Site | Google Scholar
  31. Y. Shao, J. Wang, M. Engelhard, C. Wang, and Y. Lin, “Facile and controllable electrochemical reduction of graphene oxide and its applications,” Journal of Materials Chemistry, vol. 20, no. 4, pp. 743–748, 2010. View at: Publisher Site | Google Scholar
  32. Z. Guo, Z.-y. Wang, H.-h. Wang, G.-q. Huang, and M.-m. Li, “Electrochemical sensor for isoniazid based on the glassy carbon electrode modified with reduced graphene oxide–Au nanomaterials,” Materials Science and Engineering: C, vol. 57, pp. 197–204, 2015. View at: Publisher Site | Google Scholar
  33. M. Pumera, “Graphene-based nanomaterials and their electrochemistry,” Chemical Society Reviews, vol. 39, no. 11, pp. 4146–4157, 2010. View at: Publisher Site | Google Scholar
  34. W. Yang, K. R. Ratinac, S. P. Ringer, P. Thordarson, J. J. Gooding, and F. Braet, “Carbon nanomaterials in biosensors: should you use nanotubes or graphene?” Angewandte Chemie International Edition, vol. 49, no. 12, pp. 2114–2138, 2010. View at: Publisher Site | Google Scholar
  35. M. N. L. D. Camargo, M. Santhiago, C. M. Maroneze, C. C. C. Silva, R. A. Timm, and L. T. Kubota, “Tuning the electrochemical reduction of graphene oxide: structural correlations towards the electrooxidation of nicotinamide adenine dinucleotide hydride,” Electrochimca Acta, vol. 197, pp. 194–199, 2016. View at: Publisher Site | Google Scholar
  36. Y. Yu, N. Wu, G. Peng, T. Li, L. Jiang, and H. Huang, “A simple α-ketoglutarate electrochemical biosensor based on reduced MoS2 nanoparticle-gold nanoparticle nanocomposite,” Journal of Nanoscience and Nanotechnology, vol. 18, no. 1, pp. 576–582, 2018. View at: Publisher Site | Google Scholar
  37. A. Collins, M. P. Nandakumar, E. Csöregi, and B. Mattiasson, “Monitoring of α-ketoglutarate in a fermentation process using expanded bed enzyme reactors,” Biosensors and Bioelectronics, vol. 16, no. 9–12, pp. 765–771, 2001. View at: Publisher Site | Google Scholar
  38. C.-L. Sun, H.-H. Lee, J.-M. Yang, and C.-C. Wu, “The simultaneous electrochemical detection of ascorbic acid, dopamine, and uric acid using graphene/size-selected Pt nanocomposites,” Biosensors and Bioelectronics, vol. 26, no. 8, pp. 3450–3455, 2011. View at: Publisher Site | Google Scholar
  39. L. Yang, D. Liu, J. Huang, and T. You, “Simultaneous determination of dopamine, ascorbic acid and uric acid at electrochemically reduced graphene oxide modified electrode,” Sensors and Actuators B: Chemical, vol. 193, pp. 166–172, 2014. View at: Publisher Site | Google Scholar
  40. F. Pariente, F. Tobalina, G. Moreno, L. Hernández, E. Lorenzo, and H. D. Abruña, “Mechanistic studies of the electrocatalytic oxidation of NADH and ascorbate at glassy carbon electrodes modified with electrodeposited films derived from 3,4-dihydroxybenzaldehyde,” Analytical Chemistry, vol. 69, no. 19, pp. 4065–4075, 1997. View at: Publisher Site | Google Scholar

Copyright © 2020 Gang Peng et al. This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


More related articles

 PDF Download Citation Citation
 Download other formatsMore
 Order printed copiesOrder
Views275
Downloads375
Citations

Related articles

Article of the Year Award: Outstanding research contributions of 2020, as selected by our Chief Editors. Read the winning articles.