Research Article
Development and Validation of a Hydrophilic Interaction Liquid Chromatography Tandem Mass Spectrometry Method for the Determination of Asparagine in Human Serum
Table 2
Precision and accuracy for the determination of ASN (
n
= 6).
Nominal conc. (
μ
M)
Backcalculated conc. (
μ
M)
Accuracy (%)
Precision (CV, %)
Intrabatch
Interbatch
2
1.99 ± 0.27
99.44 ± 13.52
15.02
14.74
6
5.51 ± 0.45
91.80 ± 7.48
8.54
11.21
80
79.94 ± 4.15
99.93 ± 5.18
5.96
3.89
150
145.24 ± 6.83
96.82 ± 4.55
5.03
6.02