Research Article

Low-Vacuum Quadrupole Mass Filter Using a Drift Gas

Table 2

Experimental results obtained at 20 Pa.

No.a valueq valueOperation pointPicoammeter current (nA)
Traditional method (no gas flow)Proposed method (with gas flow)

100.66Inside the stability region0.00320.1
20.10.66Inside the stability region0.00516.5
30.20.66Inside the stability region0.00512.0
40.20.2Outside the stability region0.0041.8
50.10.2Outside the stability region0.0043.6