Research Article
Simultaneous and Rapid Determination of Six Tyrosine Kinase Inhibitors in Patients with Non-Small Cell Lung Cancer Using HPLC-MS/MS
Table 3
Precision and accuracy of the LLOQ of six analytes’ calibration curves.
| Analytes | Nominal concentration (ng/mL) | Measured concentration (ng/ml) ± SD | Precision RSD (%) | Accuracy mean (%) |
| Gefitinib | 10 | 8.82 ± 0.29 | 6.31 | 96.22 | Icotinib | 50 | 50.22 ± 0.91 | 1.81 | 100.44 | Erlotinib | 50 | 49.01 ± 0.92 | 1.87 | 98.01 | Crizotinib | 10 | 10.07 ± 0.36 | 7.27 | 101.39 | Osimertinib | 10 | 9.75 ± 0.21 | 4.31 | 95.15 | Afatinib | 5 | 5.30 ± 0.31 | 5.76 | 105.98 |
|
|