Analytical Quality by Design Approach of Reverse-Phase High-Performance Liquid Chromatography of Atorvastatin: Method Development, Optimization, Validation, and the Stability-Indicated Method Table 2 Critical influencing variable and their actual and predicted values of responses.
Standard order Critical influencing variable Response Acetonitrile (%) Flow rate (ml/min) Wavelength (nm) Area (AU) Retention time Tailing factor Actual value Predicted value Actual value Predicted value Actual value Predicted value 1 50 0.35 240 1975990 1975589 2.19 2.18 1.95 1.94 2 90 0.35 240 1501486 1502578 4.18 4.19 2.42 2.41 3 50 1 240 2149670 2148578 3.23 3.22 2.12 2.12 4 90 1 240 1730395 1730796 5.33 5.34 2.47 2.48 5 50 0.68 235 2037161 2038807 2.45 2.47 2.08 2.09 6 90 0.68 235 1565069 1565223 4.84 4.84 2.51 2.52 7 50 0.68 245 2004970 2004816 2.83 2.83 2.04 2.05 8 90 0.68 245 1589253 1587607 4.62 4.6 2.44 2.43 9 70 0.35 235 1755465 1754220 2.9 2.89 2.21 2.21 10 70 1 235 1984765 1984211 4.02 4.02 2.34 2.33 11 70 0.35 245 1777250 1777804 2.98 2.99 2.14 2.15 12 70 1 245 1947774 1949019 4.04 4.05 2.26 2.26 13 70 0.68 240 1804782 1804782 3.54 3.54 2.01 2.01 14 70 0.68 240 1804782 1804782 3.54 3.54 2.02 2.01 15 70 0.68 240 1804782 1804782 3.54 3.54 2.01 2.01 16 70 0.68 240 1804782 1804782 3.54 3.54 2.01 2.01 17 70 0.68 240 1804782 1804782 3.54 3.54 2.01 2.01