Research Article

Bismuth-Remaining Cupellation Fire Assay Preconcentration Combined with Inductively Coupled Plasma Mass Spectrometry for the Simultaneous Determination of Ultratrace Au, Pt, Pd, Ru, Rh, and Ir in Geologic Samples

Table 4

Analytical performance data by the established Bi-FA ICP-MS method for Au, Pt, Pd, Ru, Rh, and Ir.

IsotopesLinear range (ng·mL−1)Linear equationR2BEC (ng·mL−1)LOD (ng·g−1)Enrichment factor

197Au0.01∼100y = 98397x + 31700.99970.0320.0252000
195Pt0.01∼100y = 70732x + 10060.99990.0140.0082000
106Pd0.01∼100y = 65527x + 124740.99950.0850.0242000
101Ru0.01∼100y = 42747x + 10660.99990.0250.0052000
103Rh0.01∼100y = 242359x + 6210.99990.0030.0022000
193Ir0.01∼100y = 198777x + 4560.99950.0020.0032000

LOD = 3standard deviation of background/slope of calibration curve, for 10 g sample. Enrichment factor was the specific value of the sample weight (10 g) before Bi-FA and the weight of Bi granule after Bi-FA (5 mg).