Research Article

Long-Term Stability of Hydromorphone in Human Plasma Frozen at −20°C for Three Years Quantified by LC-MS/MS

Table 1

Interassay accuracy and precision of the applied analytical method.

QC sampleNominal concentration (ng/mL)Mean ± SD (ng/mL)%RSD%REn

High concentration20.019.98 ± 0.331.63−0.1015
Middle concentration2.502.49 ± 0.145.45−0.5315
Low concentration0.310.35 ± 0.05214.910.915

SD: standard deviation; %RSD: relative standard deviation; %RE: relative error from nominal concentration.