Research Article

Measurement of Magnesium, Zinc, and Copper in Human Serum by Using Isotope Dilution Inductively Coupled Plasma Mass Spectrometry (ID ICP-MS)

Table 2

Repeatability of magnesium, zinc, and copper in human serum by ID ICP-MS.

Mg (mg·kg−1)Zn (mg·kg−1)Cu (mg·kg−1)
Level 1Level 2Level 3Level 1Level 2Level 3Level 1Level 2

Day 11#18.023.328.70.6250.8311.5971.061.61
18.023.328.70.6270.8341.5831.061.62
18.123.228.90.6460.8351.6101.061.62
2#18.023.328.80.6310.8341.5971.071.60
18.023.328.80.6270.8251.5941.061.62
18.023.229.00.6340.8321.5951.061.62
3#18.023.328.70.6280.8181.6071.061.61
18.023.328.70.6340.8191.5991.061.63
18.023.328.90.6250.8261.5961.061.61

Day 24#18.023.328.70.6440.8281.6011.051.60
18.023.228.70.6340.8271.6061.051.60
18.023.328.70.6330.8251.6161.051.60
5#18.023.228.70.6340.8271.5991.051.61
18.023.328.70.6320.8301.5961.061.60
18.023.328.70.6290.8261.6001.061.62
6#18.023.328.70.6280.8231.6061.061.60
18.023.228.70.6330.8301.5911.061.59
18.023.328.80.6320.8251.6021.061.60

s0.030.030.090.0060.0050.0080.0050.010

Avg18.023.328.80.6320.8281.6001.061.61

CV0.2%0.1%0.3%0.9%0.6%0.5%0.4%0.6%