Research Article

Measurement of Magnesium, Zinc, and Copper in Human Serum by Using Isotope Dilution Inductively Coupled Plasma Mass Spectrometry (ID ICP-MS)

Table 4

Sources of uncertainty in the determination.

Sources of uncertaintyMgZnCu
GBW09152GBW09152GBW09152
Value (xi)uc (xi)Value (xi)uc (xi)Value (xi)uc (xi)

Type A uncertainties
RY0.0140.0000.0340.0000.0040.000
Rz6.6670.0216.2080.0142.1070.003
RBc0.9200.0021.0410.0031.0800.004
RB0.8850.0031.0040.0031.0490.002
CB (mg·kg−1)0.50.30.090.050.0100.005
Type B uncertainties
Cz (mg·kg−1)19.3980.0031.2900.0021.0080.002
mzc (mg)3.0 × 1020.0402.6 × 1020.0405.4 × 1020.040
mYc (mg)5.5 × 1020.0402.0 × 1020.0404.0 × 1020.040
mY (mg)3.4 × 1020.040200 × 1020.0404.0 × 1020.040
mx (mg)2.0 × 1020.040300 × 1020.0404.8 × 1020.040
Combined types A and B0.250.0450.011
Degrees of freedom (Veff)888
Coverage factor (k)222
Measured value (mg·kg−1)20.571.1471.072
Expanded uncertainty (U(), k = 2) (mg·kg−1)0.20.040.08