Research Article
Measurement of Magnesium, Zinc, and Copper in Human Serum by Using Isotope Dilution Inductively Coupled Plasma Mass Spectrometry (ID ICP-MS)
Table 4
Sources of uncertainty in the determination.
| Sources of uncertainty | Mg | Zn | Cu | GBW09152 | GBW09152 | GBW09152 | Value (xi) | uc (xi) | Value (xi) | uc (xi) | Value (xi) | uc (xi) |
| Type A uncertainties | RY | 0.014 | 0.000 | 0.034 | 0.000 | 0.004 | 0.000 | Rz | 6.667 | 0.021 | 6.208 | 0.014 | 2.107 | 0.003 | RBc | 0.920 | 0.002 | 1.041 | 0.003 | 1.080 | 0.004 | RB | 0.885 | 0.003 | 1.004 | 0.003 | 1.049 | 0.002 | CB (mg·kg−1) | 0.5 | 0.3 | 0.09 | 0.05 | 0.010 | 0.005 | Type B uncertainties | Cz (mg·kg−1) | 19.398 | 0.003 | 1.290 | 0.002 | 1.008 | 0.002 | mzc (mg) | 3.0 × 102 | 0.040 | 2.6 × 102 | 0.040 | 5.4 × 102 | 0.040 | mYc (mg) | 5.5 × 102 | 0.040 | 2.0 × 102 | 0.040 | 4.0 × 102 | 0.040 | mY (mg) | 3.4 × 102 | 0.040 | 200 × 102 | 0.040 | 4.0 × 102 | 0.040 | mx (mg) | 2.0 × 102 | 0.040 | 300 × 102 | 0.040 | 4.8 × 102 | 0.040 | Combined types A and B | — | 0.25 | — | 0.045 | — | 0.011 | Degrees of freedom (Veff) | | 8 | | 8 | | 8 | Coverage factor (k) | — | 2 | — | 2 | — | 2 | Measured value (mg·kg−1) | 20.57 | — | 1.147 | — | 1.072 | — | Expanded uncertainty (U(), k = 2) (mg·kg−1) | — | 0.2 | — | 0.04 | — | 0.08 |
|
|