Research Article

Comparing Jaffe and Enzymatic Methods for Creatinine Measurement at Various Icterus Levels and Their Impacts on Liver Transplant Allocation

Table 1

Characteristics of methods used in the study.

AnalytesMethodReagentWavelength (nm)Analytical sensitivity (mg/dL)Linearity limit (mg/dL)Limit of icterus interference, bilirubin (mg/dL)

CreatinineJaffeBIOREX5000.2204
CreatinineEnzymatic (creatinine deiminase)BIOREX3400.22015
BilirubinJendrassik- GrofBIOREX5460.125

IBM SPSS (version 25.0) was used to analyze all the data. Quantitative variables were expressed as mean ± SD and/or median (range). Significance testing was 2-sided and set to less than 0.05. The Wilcoxon signed-rank test was used for a nonparametric comparison between paired Cr values and paired MELD scores. The Mann–Whitney U test was used to determine how the mean values differed.