Research Article

Exploiting of Green Synthesized Metal Oxide Nanoparticles in the Potentiometric Determination of Metformin Hydrochloride in Pharmaceutical Products

Table 5

Intraday and interday assays of MTF solution using MTF-PT-Al2O3NP-modified and MTF-PT-NiONP-modified sensors.

MTF-PT-Al2O3NP-modified sensorMTF-PT-NiONP-modified sensor
Test sampleFound% recoveryTest sampleFound% recovery

Intraday1010.00100.00109.9999.90
87.9699.5087.9599.38
65.9999.9365.9899.67

Mean ± SD99.81 ± 0.2799.65 ± 0.26

n33

SE%∗∗0.160.15

RSD%0.270.26

Interday109.9999.90109.9599.50
87.9499.2587.9499.25
66.00100.0065.9899.67

Mean ± SD99.72 ± 0.4199.47 ± 0.21

n33

SE%∗∗0.240.12

RSD%0.410.21

SE (% Error) = %RSD/√n.