Phantom composition in weight percent, target, and measured dielectric parameters at 1 GHz and 25°C, immediately before the measurement in the MWI system.

HEADHEM

Isopropyl alcohol (wt%)50.4834.00
Deionized water (wt%)48.3564.67
NaCl (wt%)1.161.33

Target (-)39.6051.48
Measured (-)39.8051.37
U () (%)1.72.1

Target σ (S/m)0.941.220
Measured σ (S/m)0.941.224
U () (%)2.72.7

U: expanded uncertainty of type B, HEAD: head phantom, HEM: hemorrhagic phantom.