Anatomically and Dielectrically Realistic 2.5D 5-Layer Reconfigurable Head Phantom for Testing Microwave Stroke Detection and Classification

Table 4

Dielectric properties at 1 GHz of the different tissue-equivalent layers of the human head phantom for stroke detection. The measured data are supplemented with extended uncertainties (k = 2).

Tissue layerεr (−) humanεr (−) measuredσ (S/m) humanσ (S/m) measured

Scalp31.0634.16 ± 1.330.640.77 ± 0.08
Skull11.9712.17 ± 0.530.090.11 ± 0.01
CSF68.4482.81 ± 2.462.462.99 ± 0.04
Brain46.5745.33 ± 0.780.970.98 ± 0.01
Ischemic stroke37.2636.43 ± 0.660.780.87 ± 0.01
Hemorrhagic stroke61.0855.78 ± 2.181.591.80 ± 0.09