Research Article
Design and Experimental Assessment of a 2D Microwave Imaging System for Brain Stroke Monitoring
Table 1
Case 1 evaluated metrics.
| Green’s function of the homogeneous coupling medium | | Raw data at 1-1.5 GHz | TG data at 1-1.5 GHz | De-emb. data at 1-1.5 GHz | TG de-emb. data at 1-1.5 GHz | Dim | 0.83 | 1.31 | 0.86 | 1.31 | Loc | 3.03 mm | 4.71 mm | 2.5 mm | 6.9 mm | SLL | 0.88 dB | -2.38 dB | 0.73 dB | -3.1 dB | | Raw data at 1.5-1.75 GHz | TG data at 1.5-1.75 GHz | De-emb. data at 1.5-1.75 GHz | TG de-emb. data at 1.5-1.75 GHz | Dim | 1.33 | 1.1 | 1.14 | 1 | Loc | 2.51 mm | 2 mm | 3 mm | 4.2 mm | SLL | -1.41 dB | -4.4 dB | -1.94 dB | -4 dB |
| Green’s function of the homogeneous coupling medium with inside the brain phantom | | Raw data at 1.5-1.75 GHz | TG data at 1.5-1.75 GHz | De-emb. data at 1.5-1.75 GHz | TG de-emb. data at 1.5-1.75 GHz | Dim | 1.17 | 0.53 | 1.08 | 0.58 | Loc | 2.5 mm | 1.8 mm | 1.8 mm | 1.5 mm | SLL | -0.5 dB | -4.44 dB | -1.83 dB | -4.73 dB |
|
|