Research Article

Design and Experimental Assessment of a 2D Microwave Imaging System for Brain Stroke Monitoring

Table 1

Case 1 evaluated metrics.

Green’s function of the homogeneous coupling medium
Raw data at 1-1.5 GHzTG data at 1-1.5 GHzDe-emb. data at 1-1.5 GHzTG de-emb. data at 1-1.5 GHz
Dim0.831.310.861.31
Loc3.03 mm4.71 mm2.5 mm6.9 mm
SLL0.88 dB-2.38 dB0.73 dB-3.1 dB
Raw data at 1.5-1.75 GHzTG data at 1.5-1.75 GHzDe-emb. data at 1.5-1.75 GHzTG de-emb. data at 1.5-1.75 GHz
Dim1.331.11.141
Loc2.51 mm2 mm3 mm4.2 mm
SLL-1.41 dB-4.4 dB-1.94 dB-4 dB

Green’s function of the homogeneous coupling medium with inside the brain phantom
Raw data at 1.5-1.75 GHzTG data at 1.5-1.75 GHzDe-emb. data at 1.5-1.75 GHzTG de-emb. data at 1.5-1.75 GHz
Dim1.170.531.080.58
Loc2.5 mm1.8 mm1.8 mm1.5 mm
SLL-0.5 dB-4.44 dB-1.83 dB-4.73 dB