Effect of Edge-Preserving Adaptive Image Filter on Low-Contrast Detectability in CT Systems: Application of ROC Analysis

Table 1

95% confidence intervals of the differences in average Az values and P-values obtained using MRMC analysis for low-contrast detection task for each target size (2, 3, 4, 5 mm) using images with and without QDS filter. These results also list the tube current used for each condition (target size, with or without QDS filter) to obtain a mean Az value that just exceeded 0.7. In all cases, “0” was included in the 95% confidence intervals and the P-values were large.

Target diameterQDSTube currentMean Az valueDifference95% confidence intervalP-value

2 mmWithout320 mA0.709 −0.051 −0.176 0.074 0.410
With200 mA0.760
3 mmWithout180 mA0.706 0.002 −0.074 0.078 0.911
With120 mA0.704
4 mmWithout140 mA0.730 −0.029 −0.133 0.075 0.566
With90 mA0.759
5 mmWithout100 mA0.717 −0.004 −0.080 0.072 0.880
With80 mA0.721