Evaluation of Sterilisation Techniques for Regenerative Medicine Scaffolds Fabricated with Polyurethane Nonbiodegradable and Bioabsorbable Nanocomposite Materials

Table 1

Mechanical Properties of POSS-PCU and POSS-PCL after sterilisation processes. Young’s modulus, maximum stress, elongation at break, and thickness of control POSS-PCL and POSS-PCU and samples subjected to ethanol, bleach (SDIC), plasma, ethylene oxide, UV radiation, antibiotic/antimycotic treatment, and gamma irradiation.

Sterilisation MethodControlBleach (SDIC)EthanolGammaEthylene OxideUVAntibiotic/AntimycoticPlasmaAutoclaving

Young’s Modulus (MPa)POSS-PCL0.18 ± 0.00.29 ± 0.020.20 ± 0.020.32 ± 0.040.40 ± 0.110.35 ± 0.020.25 ± 0.030.30 ± 0.05N/A
POSS-PCU0.55 ± 0.040.57 ± 0.020.56 ± 0.020.56 ± 0.040.55 ± 0.020.55 ± 0.020.53 ± 0.010.55 ± 0.030.54 ± 0.02
Maximum Stress (MPa)POSS-PCL0.56 ± 0.080.61 ± 0.100.31 ± 0.111.17 ± 0.151.71 ± 0.191.45 ± 0.041.10 ± 0.140.80 ± 0.26N/A
POSS-PCU0.83 ± 0.030.82 ± 0.060.85 ± 0.030.83 ± 0.010.84 ± 0.010.83 ± 0.010.81 ± 0.120.83 ± 0.020.82 ± 0.04
Elongation at break (%)POSS-PCL589.6 ± 26.2549.7 ± 68293.0 ± 61462.2 ± 16471.6 ± 39387.5 ± 22469.5 ± 11500.1 ± 30N/A
POSS-PCU283.3 ± 9.57273.6 ± 10.01279.2 ± 1.74288.0 ± 3.05288.6 ± 1.46269.3 ± 8.55273.8 ± 9.01281.1 ± 4.12280.5 ± 7.88
Thickness (mm)POSS-PCL2.0 ± 01.43 ± 0.151.525 ± 0.131.6 ± 0.221.2 ± 0.251.38 ± 0.051.55 ± 0.172.433 ± 0.12N/A
POSS-PCU0.75 ± 0.040.77 ± 0.020.84 ± 0.020.77 ± 0.050.76 ± 0.020.77 ± 0.020.75 ± 0.030.76 ± 0.020.74 ± 0.06