International Journal of Dentistry / 2009 / Article / Tab 5 / Research Article
The Efficiency of Operating Microscope Compared with Unaided Visual Examination, Conventional and Digital Intraoral Radiography for Proximal Caries Detection Table 5 Pairwise comparisons of
values.
Pairwise Difference between area
Std. error (a) P value Asymptotic 95% confidence interval Lower bound Upper bound
1st Observer
Operating microscope-unaided
visual examination 0.000 0.051 .996
0.0990.099
Operating microscope-film radiography 0.150 0.072 .036 0.010 0.291
Operating microscope-RVG 0.143 0.072 .048 0.001 0.285
Unaided visual
examination-film radiography 0.150 0.072 .038 0.009 0.291
Unaided visual examination-RVG 0.143 0.073 .050 0.000 0.285
Insight-RVG 0.007 0.054 .896
0.0990.113
3rd Observer
Operating microscope-unaided
visual examination 0.001 0.036 .984
0.0700.071
Operating microscope-film radiography 0.240 0.078 .002 0.087 0.393
Operating microscope-RVG 0.226 0.078 .004 0.074 0.379
Unaided visual
examination-film radiography 0.241 0.078 .002 0.088 0.394
Unaided visual examination-RVG 0.227 0.078 .003 0.075 0.380
Film radiography-RVG 0.014 0.047 .772
0.0780.106