Factors Influencing Resonance Frequency Analysis (RFA) Measurements and 5-Year Survival of Neoss Dental Implants

Table 10

Implant survival below and above at different threshold values at implant insertion and loading.

VariableGroupsImplants placedImplant failuresCSR(1) (%)Odds ratio95% C.I. value
OverallEarly/late

ISQ at insertion<603921/194.91.900.4–8.60.314(1)
≥60616176/1197.21

ISQ at loading<60600/0100N/AN/A1.0(1)
≥60523120/1297.7

ISQ at insertion<658351/494.02.560.9–7.30.080(1)
≥65572146/897.61

ISQ at loading<654220/295.22.390.5–11.20.245(1)
≥65487100/1097.91

ISQ at insertion<70151102/893.43.901.6–9.70.004(1)
≥7050495/498.21

ISQ at loading<7010170/793.16.301.9–20.20.003(1)
≥7042850/598.81

ISQ at insertion<75293166/1094.56.902.0–23.80.001(1)
≥7536231/299.21

ISQ at loading<75207110/1194.717.92.3–142.9<0.001(1)
≥7532210/199.71

CSR: cumulative survival rate; (1)Fisher’s exact test.