Research Article

Clinical Evaluation of Subcutaneous Lactate Measurement in Patients after Major Cardiac Surgery

Table 3

Clinical evaluation of 6-hour lactate trend analysis from subcutaneous adipose tissue ( satLAC) versus arterial blood ( artLAC). Lactate measurements are stratified to positive, stable and negative lactate trends as indicated in Figure 1(b).

6-hourly paired lactate trendsAcceptable measurement %Unacceptable falsely high %Unacceptable falsely low %

Positive trend ( 20%)61.911.127.0
Negative trend ( −20%)66.727.06.3
No change (±20%)66.420.912.7

Total trend analysis65.319.914.8