Research Article
Estimation of True Serum Thyroglobulin Concentration Using Simultaneous Measurement of Serum Antithyroglobulin Antibody
Table 3
Decline of measured Tg value by TgAb in samples produced using Tg and TgAb standard solutions.
| Expected Tg concentration (ng/mL) | Measured Tg concentration (ng/mL) | under various TgAb concentrations | 10.0 U/mL | 30.0 U/mL | 100.0 U/mL | 300.0 U/mL | 1000.0 U/mL |
| 2.0 | 5.1 ± 0.1 | 4.0 ± 0.2 | 3.4 ± 0.1 | 2.6 ± 0.1 | 1.8 ± 0.1 | 10.0 | 13.0 ± 0.5 | 11.8 ± 0.4 | 10.9 ± 0.7 | 10.0 ± 0.3 | 6.2 ± 0.3 | 50.0 | 52.8 ± 1.8 | 47.1 ± 1.0 | 41.4 ± 1.1 | 40.3 ± 1.2 | 30.6 ± 0.5 | 125.0 | 134.1 ± 3.2 | 128.7 ± 2.0 | 90.0 ± 1.0 | 95.7 ± 2.5 | 75.5 ± 3.0 |
|
|
Values are expressed as mean ± SD.
|