Research Article

Blood Glucose Control Using a Novel Continuous Blood Glucose Monitor and Repetitive Intravenous Insulin Boluses: Exploiting Natural Insulin Pulsatility as a Principle for a Future Artificial Pancreas

Table 1

The seven outcome variables from the four animals; BGL0: the blood glucose level at the time of the first insulin bolus; BGLlow: the lowest blood glucose level recording during the experiment; : time to establish glycemic control from the first insulin bolus until the blood glucose level went below 6.0 mmol/L; RD: the rate of blood glucose level decrease during ; : time with glycemic control from blood glucose level went below 6.0 mmol/L until end of experiment; : total time with blood glucose levels below 4.5 mmol/L during ; : percentage time with BGL in the correct range between 4.5 and 6.0 mmol/L during .

AnimalBGL0 (mmol/L)BGLlow (mmol/L) (min)RD ((mmol/L)/min) (min) (min) (%)

17.533.81680.0251865173
27.464.14210.0731462086
39.334.44630.064238399
414.064.041210.0771282283