Research Article

Single-Walled-Carbon-Nanotube-Modified Pyrolytic Graphite Electrode Used as a Simple Sensor for the Determination of Salbutamol in Urine

Table 3

A comparison of voltammetric response of SWNT/EPPGE with previously reported methods for the determination of salbutamol.

Sr. no.ElectrodeConcentration range (μM)Limit of detection (μM)Sensitivity (μA/μM)Analytical utility real samplesRef. no.

(1)Pt electrode100–100080.000.012No[22]
GCE20–100010.000.066No[22]
(2)NGITO0.20–8.350.310.055Yes[23]
(3)GCE0.80–80.000.20No[24]
(4)MWNT/GCE0.80–10.000.200.620No[25]
(5)C60/GCE0.42–8.350.170.048Yes[26]
(6)GN/GCE5–900.10No[27]
(7)SWNT/EPPGE0.20–10.450.0180.520YesProposed method