Research Article

Enantioselective Potentiometric Membrane Electrodes Based on Antibiotics for the Determination of L- and D-Glyceric Acids

Table 1

Response characteristics of enantioselective, potentiometric membrane electrodes for L- and D-glyceric acidsa.

Parameters
EPME based onSlopeIntercept, Linear rangeDetection limit
(mV/decade of concentration)(mV)(moL/L)(moL/L)

Vancomycin58.6574.610−9–10−7
Teicoplanin50.0206.010−4–10−2

aAll measurements were made at 25°C; all values are the average of ten determinations.