Research Article

Voltammetric Determination of Uric Acid in Clinical Serum Samples Using DMF Modified Screen Printed Carbon Electrodes

Table 1

Comparison of the electrochemical performance of current sensor with reported ones.

Electrode Modifier usedMethod LDR LOD (M)Reference

CPENiHCFDPV2 - 12 μM1.8 x 10−7 [1]
SPCENiHCFCV50 μ - 1.5 mM5.5 x 10−6[15]
GCESWCNTsDPV100 nM to 3400 μM30 nM[21]
SPCEDMFDPV5-100 μM and 100 - 500 μM1.9 x10−7This work