Research Article

Evaluation of Handheld Assays for the Detection of Ricin and Staphylococcal Enterotoxin B in Disinfected Waters

Table 1

LODsa or ricin and SEB in multiple water matrices using handheld assays.

Manufacturer 1Manufacturer 2
1-line2-line1-line2-line2-dot

Ricin LODs

Tap1.17.11.21.35.9
Tap + Cl1.816.71.01.31.3
RO + Cl3.356.92.22.54.9
RO + Br30003000300030003000

SEB LODs

Tap3.112.50.91.11.9
Tap + Cl25005000250050005000
RO + Cl250010000250050005000
RO + Br25005000250050005000

aAll LODs reported are in ng/mL.
Concentrations were tested a minimum of two independent times with a minimum of three handheld assays per concentration.