Research Article

Comparison of Laboratory Diagnosis of Urinary Tract Infections Based on Leukocyte and Bacterial Parameters Using Standardized Microscopic and Flow Cytometry Methods

Table 3

Cross-tabulation and a kappa value of Sysmex UF-5000 and Shih-Yung (bacterial parameter).

BACT (Sysmex UF-5000)BACT (Shih-Yung)Total
None, n (%)Rare, n (%)Few, n (%)Moderate, n (%)Many, n (%)

None1 (100)0 (0)0 (0)0 (0)0 (0)1 (100)
Rare0 (0)22 (100)0 (0)0 (0)0 (0)22 (100)
Few0 (0)0 (0)6 (100)0 (0)0 (0)6 (100)
Moderate0 (0)0 (0)0 (0)2 (66.7)1(33.3)3 (100)
Many0 (0)0 (0)0 (0)2 (2.9)66 (97.1)68 (100)
Kappa0.939
value<0.001